Conmed Sabre 2400 Electrosurgical Unit

Powers on and no further testing was done.

Product Enquiry
Category: Tag:

Description

Features independent Monopolar outputs, enhanced bipolar performance. It is safe, convenient, cost-effective and can be programmed for standardization.

Conmed Sabre 2400 Specifications:

General

Pure Cut
Max Power: 300 W.
Max. P-P Voltage: 2000<./li>
Load (ohms): 300.
Crest Factor: 1.8.
Blend Cut
Max Power: 180 W.
Max. P-P Voltage: 2200.
Load (ohms): 300.
Crest Factor: 2.5.
Coag
Max Power: 120 W.
Max. P-P Voltage: 9000.
Load (ohms): 500.
Crest Factor: 7.0.
Bipolar
Max Power: 50 W.
Max. P-P Voltage: 280.
Load (ohms): 50.
Crest Factor: 1.5 – 12.1.

Dispersive Electrode Monitor

With a dual foil dispersive electrode, the Aspen Return Monitor will alarm above 150 ohms, below 10 ohms, and with a 20% increase in resistance above set point. With a single foil dispersive electrode, a two-wire continuity detector alarms at 10 ohms and above.

Operating Frequency

Cut: 416.7 kHz ± 0.01%.
Coag: 416.7 kHz ± 0.01%.

Pulse Repetition Frequency

Blend: 20.0 kHz ± 0.01%.
Coag: 25.0 kHz ± 0.01%.
Bipolar: 20.0 kHz ± 0.01%.

Input Voltage

104-127 VRMS, 50/60 Hz, calibrated at 120V.

Input Current

Pure cut at 300: 7.0A max.

Line Frequency Leakage.

Patient connections to neutral: < 25 μA.
Chassis to neutral: < 25 μA typical.

RF Leakage Current

Less than 100 mA (200 ohms to ground).

Power Decrease

1% typical. Pure cut set to 150 watts for 20 seconds.

Weight

32 lbs (14.5 kg).

Dimensions

6.75″H x 14.25″W x 21.5″D (17.1 x 36.2 x 54.6 cm).