Bear Cub 750 VS Infant Ventilator

Bear Cub® 750 Infant Ventilator is designed to accommodate a range of patients from the small neonate to the 30kg pediatric patient. The 750 incorporates Integrated Synchronized Ventilation, Tidal Volume Monitoring and Independent Base & Inspiratory Flows for clinical flexibility in a compact design. Volume Limit™ provides the clinician with an additional guidance tool during pressure ventilation as it enhances volume management and establishes an upper boundary for deliverable tidal volumes during mechanical breaths.

Product Enquiry
Category: Tag:

Description

FEATURES:

Integrated Synchronized Ventilation
Tidal Volume Monitoring
Independent Base & Inspiratory Flows
Internal Battery Back-Up
Exclusive Volume Limit
Assist Control
Flow Cycled AC
SIMV/IMV
Flow Cycled SIMV
CPAP
SIMV with Pressure Support

SPECIFICATIONS:

Controls

Mode: AC, Flow cycle AC, SIMV/IMV, Flow Cycled SIMV, SIMV/PSV, PSV, CPAP
Ventilator Rate: 1.0 to 150 BPM
Inspiratory Time: 0.1 to 3.0 sec.
Inspiratory Flow: 1.0 to 30 LPM
Base Flow: 1.0 to 30 LPM
Volume Limit: 5.0 to 300 ml
PEEP: 0 to 30 cmH2O
Inspiratory Pressure: 0 to 72 cmH2O
Manual Breath: x One
Assist Sensitivity: 0.2 to 5.0 LPM; Minimum to Maximum bar
Over Pressure Relief: 15 to 75 cmH2O
% O2: 21 to 100%
Alarm Loudness: Min 60 to Max. 75 dbA
Audible Off Volume Limit: LED On/Off

Alarms

High Breath Rate: 3.0 to 255 BPM
Low PEEP/CPAP: -5.0 to 30 cmH2O
Low Inspiratory Pressure (vs): 1.0 to 65 cmH2O
Low Inspiratory Pressure (psv): 5 to 41 cmH2O (derived from High Pressure Limit and Low PEEP/CPAP Alarm Settings
Patient Circuit: LED On/Off
Failed to Cycle: LED On/Off
Low Gas Supply: LED On/Off
Apnea Alarm: 5, 10, 20, 30 sec
Settings Incompatible: LED On/Off
Pressure Settings Incompatible: LED On/Off
Prolonged InspiratoryPressure: LED On/Off
Flow Sensor: LED On/Off
Low Battery: LED On/Off
Alarm Silence: 60 sec
Visual Reset: Push Button
High Pressure Limit: 10 to 75 cmH2O
Low Minute Volume: Off to 99 lpm

Monitors

PEEP: 0 to 30 cmH2O
Breath Rate: 0 to 255 BPM
Breath Type (Patient Initiated):LED
Minute Volume: 0 to 30.0 LPM
Tidal Volume (Exhaled): 0 to 500 ml
% Tube Leak: 0 to 100%
Inspiratory Time: 0.1 to 3.10 sec Expiratory Time 9 to 99.9 sec
I:E Ratio: 9.9:1 to 1:9.9
Peak Inspiratory Pressure: 0 to 99 cmH2O
Mean Airway Pressure: 0 to 75 cmH2O
Air Pressure: 0 to 100 psig; 0 to 99 psig
O2 Pressure: 0 to 100 psig; 0 to 99 psig
Proximal Airway Pressure: -10 to 100 cmH2O
Hour meter: 0 to 99,999 hours
Test: Push Button
Line Power: Green/Red LED
Battery: LED On/Off
Inspired Tidal Volume: 0 to 500 ml

Limits to Ventilation (Fixed)

Minimum Expiratory Time: 0.15 seconds
Maximum Inspiratory Time: 3:10 seconds
Maximum settable I:E Ratio: 4:1

TOLERANCES:

Monitors

Breath Rate: ‚± 1 BPM or ‚± 20 msec, whichever is greater
Inspiratory Time: ‚± 0.02 sec
Expiratory Time: ‚± 0.02 sec
I:E Ratio: ‚± 0.1 or ‚± 20 msec (on the calculation based on the monitored inspiratory and expiratory times), whichever is greater
Airway Pressure (Continuous):
‚± 1 cmH2O (-10 to 20 cmH2O)
‚± 2 cmH2O (20 to 65 cmH2O)
‚± 3 cmH2O (65 to 100 cmH2O)
Peak Inspiratory Pressure: ‚± 2 cmH2O or ‚± 5%, whichever is greater
Mean Airway Pressure: ‚± 2 cmH2O or ‚± 3%, whichever is greater
PEEP: ‚± 2 cmH2O or ‚± 5%, whichever is greater
Air/O2 Pressure: ‚± 5 psig
Minute Volume Accuracy: Based on the Tidal Volume and Breath Rate Monitors
Inspiratory Tidal Volume: ‚± 1 ml or ‚± 10%, whichever is greater
Expiratory Tidal Volume: ‚± 1 ml or ‚± 10%, whichever is greater
% Leak: ‚± 1 ml or ‚± 2%, whichever is greater
(when compared to the Inspired and Expired Tidal Volume Monitors)
Hour Meter: ‚± 2% of reading

Alarms/Limits

Low PEEP/CPAP: ‚±2 cmH2O
High Breath Rate: ‚±1 BPM
Low Minute Volume:
‚±10 ml/min (0 to 99 ml/min)
‚±0.1 L/min (0.1 to 9.9 L/min)
High Pressure Limit: ‚±4 cmH2O
Prolonged Inspiratory Pressure: ‚±2 cmH2O
Volume Limit: ‚±2 ml or ‚± 10% of setting, whichever is greater
Apnea: ‚±1 sec
Maximum (Pop-off) Pressure: ‚±4 cmH2O

Controls

Ventilator Rate: ‚±1 BPM or ‚±20 msec (applied to the breath interval), whichever is greater
Inspiratory Rate:
‚±0.020 sec (0.10 to 0.50 sec)
‚±0.025 sec (0.50 to 3.00 sec)
Assist Sensitivity: Not a calibrated scale
PEEP/CPAP and Inspiratory: ‚±1 cmH2O or ‚± 5%, whichever is greater (breath to breath)
Base/Inspiratory Flow: ‚±0.5 LPM or ‚±10% of setting, whichever is greater (0 to 60 cmH2O)
Oxygen %: ‚± 3%
Degree of Safety in Presence of Flammable Anesthetics: None
Mode of Operation: Continuous

Breathing Circuit Parameters

Inspiratory Resistance: 0.42 cmH2o/l/min
Expiratory Resistance: 0.08 cmH2o/l/min
Compliance: 1.34 ml/cmH2o
Internal Volume: 1370 ml

Outputs

Digital: RS-232 Bi-Directional
Analog: See Manual
Proximal Pressure: -10 to 100 cmH2O, 1 cmH2)/25 mV
Proximal Inspiratory/Expiratory Flow: -40 (expiratory) to 40 (inspiratory) LPM, 1 LPM/50 mV
Machine Delivered Flow: 0 to 30 lpm, 1 LPM/50 mV
Breath Phase: 0.5V Logic Signal
Remote Nurse Call: 0.5 amps max at 24 vdc
Auxiliary Blended GasOutlet: 7 to 17 psig, 0 to 8 LPM

Inputs

Electrical

Voltage:
100 V – 80 to 110 VAC
120V – 96 to 132 VAC
220V – 176 to 242 VAC
240V – 192 to 264 VAC
Current:
100V – 1.0 A maximum
120V – 1.0 A maximum
220V – 0.5 A maximum
240V – 0.5 A maximum
Frequency:
100V – 50.60 Hz
120V – 50/60 Hz
220V – 50/60 Hz
240V – 50/60 Hz
Fuses:
100/120V – T o.5 A, 5 x 20 mm
230/240V – T 0.5 A, 5 x 20 mm
(For 220V operation, set the power entry module seclector switch to 230V setting)

Pneumatic:

Oxygen and Air:
30 to 80 psig, 50 LPM Supply Ratings
KPa (206 to 551)

Physical Dimension and Shipping Information:

Ventilator Weight: 12 kg / 13.6 kg
Graphics Display Weight: 2.5 kg
Pedestal Stand Weight: 4.1 kgCompressor Weight: 50 kg

Ventilator Dimensions: 13.5″W x 10″D x 11″H
Graphics Display Dimensions: 13″W x 2.5″D x 9.5″H
Pedestal Stand Dimensions: 24″ x 40.5″H
Compressor Dimensions: 22″W x 21.5″D x 36″H

Shipping Weight Including Ventilator: 17 kg; 18.2 kg
Graphics Display: 5 kg
Pedestal Stand: 5.1 kg
Compressor: 9.1 kg

Shipping Dimensions
Ventilator: 18″W x 18″D x 19″H
Graphics Display: 16″W x 13″D x 11″H
Pedestal Stand: 25″W x 5″D x 45″H
Compressor: 26″W x 22″D x 40″H

Environmental Specifications:

Temperature

Storage and Shipping: -40 to 70 C
Checkout and Operating: 10 to 40 C

Altitude

Checkout, Operating, Transport and Storage 0 to 3,000 meters (14.7 to 10.5 PSIA/760 to 543 mm Hg)

Humidity

Storage and Shipping: 0 to 99% Relative Humidity – Non-condensing
Checkout and Operating: 0 to 95% Relative Humidity – Non-condensing